Conductivity sensor LMA05
with current/voltage output, 4-electrode technology with Pt100
- Conductivity sensor with 4-electrode technology
- Measuring range programmable ex works
- 0…200µS/…2000µS/
…20mS/…200mS/…800mS/cm - with ISO measuring amplifier and output:
- 0(4)…20mA
- 0(2)…10V DC
- Output is galvanically isolated
- insensitive to contamination
- for plastic union KV50 2-1/4" or immersion fitting TA32 possible
- Made in Germany
ab €592.00 net
Advantages
Insensitive to contamination thanks to 4-electrode technology:
Contamination of the electrode with insulating substances (oils etc.) changes the cell constant and falsifies the measurement results.
With 4-electrode sensors, an additional currentless electric field measurement is carried out via sense measurement. The sensor is therefore less sensitive to contamination.
Integrated output electronics:
Output current or voltage is generated in the sensor. You save the cost of a transmitter or display unit; current or voltage can be fed directly into a PLC or similar. Even long cable runs can no longer falsify the measurement result.
Fully analogue:
The measuring amplifier is equipped exclusively with analogue components. There is no falsification of the measurement results due to analogue/digital conversion.
Programming ex works
The device is programmed to your requirements on delivery.
Please specify when ordering:
- Measuring range
- Temperature compensation
- Analogue output.
Dimensions and pin assignment
| green = current/voltage output white = UB+ brown = UB ground |
Technical data
| Programmable measuring ranges: | 0…200/ 2000µS/…20/…200/…800mS/cm | ||||||
| Measuring accuracy [%]: | ±2 of full scale | ||||||
| Degree of protection: | IP 65 | ||||||
| Temperature compensation: |
| ||||||
| Measuring surfaces: | Electrographite / titanium / platinum | ||||||
| Cell shaft: | PP, up to max. +50°C | ||||||
| Seal: | O-ring Viton | ||||||
| Output: | 0(4)…20mA DC / 0(2)…10V DC (with galvanic isolation) | ||||||
| Pressure resistance [bar]: | 16 up to 60°C | ||||||
| Current consumption [mA]: | max. 115 | ||||||
| Fixed cable connection: | 5m LIYY 3x0.34mm² use shielded cable for lengths >5m. | ||||||
| Operating voltage [VDC]: | 18…28 | ||||||
| Current consumption [mA] | max. 115 | ||||||
| Installation: | Immersion fitting TA32 KV50 2-1/4" |
Items & prices
| Art.-Nr. | Description | Net price | Order / enquire |
|---|---|---|---|
| 1-0022-0307 | Conductivity sensor LMA05Active measuring cell with measuring amplifier4-electrode technology, insensitive to contaminationfor plastic union KV50 2-1/4" and/or immersion fitting TA32measuring range programmable ex works 0…200µS/…2000µS/…20mS/…200mS/…800mS/cmOutput 0/4…20mA, 0/2…10VDCMaterial: PP, Viton, titanium, electrographite, platinum | €592.00 |
All prices in euros net plus statutory VAT and shipping costs. Errors and changes excepted. Our terms and conditions apply: AGB.