FAQ: pH measurement

What does the pH value indicate?

A high concentration of hydrogen ions in water means acidic water, a low concentration means alkaline water. At a concentration of 10-7 = pH 7 the reaction is neutral.
pH values below pH 7 indicate acidic liquids, pH values above pH 7 alkaline liquids. Due to the logarithmic unit, each whole number on the pH scale means a tenfold change in acidity (degree of alkalinity). pH measurement is used today in large parts of industry, medicine, pharmacy and research.

How does an electronic pH meter work?

A glass probe is immersed in the solution to be tested. The probe is connected via a cable to a battery-powered meter. The electrical voltage generated by the probe depends on the pH value. A highly sensitive measuring amplifier processes this measuring signal and feeds it to the digital display. The accuracy of the instrument can be checked at any time using two precisely adjusted calibration solutions (buffer solutions).

How long does a pH measurement take?

The time a pH sensor needs to deliver a stable reading ranges from a few seconds to several minutes. It depends on the design and age of the sensor as well as the consistency and temperature of the measuring solution.

Electrode, measuring chain or sensor. Which is correct?

The terms electrode, measuring chain or combination electrode are often used synonymously.
Usually, combination electrodes are meant, i.e. an electrode combined with a reference electrode, whereby the reference electrode also refers to the reference system.
We consistently use the term Sensor.

What exactly is a combination electrode?

Measuring chains for measuring electrochemical potentials always consist of a measuring electrode and a reference electrode. The measuring electrode outputs a voltage in relation to the concentration. The voltage of the reference electrode, on the other hand, is constant and is not influenced by the medium.
Combination electrodes form a single unit of glass and reference electrode. The reference electrode surrounds the glass electrode concentrically.

Reference electrode system:
Reference electrode systems are mainly equipped with a silver/silver chloride reference system. Reference systems with liquid electrolyte should always contain sufficient liquid; if necessary, they are topped up with 3 mol potassium chloride solution. During measurement, the filling opening should be exposed for pressure equalisation. The liquid column inside the electrode should rise a few centimetres above the surface of the measuring solution to create a hydrostatic overpressure, which ensures a constant flow of KCl solution through the diaphragm out into the measuring solution. This ensures constant self-cleaning of the diaphragm and also prevents possible electrode poisons from diffusing into the electrode and rendering the lead-off system unusable. Reference electrodes with gelled reference electrolyte cannot be refilled.
Diaphragms:
Reference electrodes are connected to the measuring solution via a diaphragm. This diaphragm is porous and permeable to liquid and usually consists of sintered ceramic.
The pH readings obtained are obviously incorrect. What are possible causes?

If obviously incorrect measurements occur, various causes are possible.

  • Reference in stationary medium, current measurement in flowing medium.
    If the reference measurement was taken in a stationary medium but the current measurement is taken in a flowing medium, the different readings can be explained by mixing.
    In a stationary medium, stratification occurs in the medium; the pH values change depending on the layer.
    In a flowing medium, the medium is mixed and the pH value changes.
  • Incorrect configuration of the display unit.
    There may be an incorrect configuration, e.g. temperature compensation is not activated.
Can the pH transmitter be operated with a fixed resistor instead of a Pt100 temperature sensor?

Ja.
This is possible if the pH value of a medium at constant temperature is measured. In this case, however, a fixed resistor must always be connected to terminals 5 and 6.
Depending on the temperature of the medium, use the following resistance values:

0 °C: 100 Ohm
10 °C: 103.9 Ohm
20 °C: 107.79 Ohm
30 °C: 111.67 Ohm
40 °C: 115.54 Ohm
50 °C: 119.40 Ohm
60 °C: 123.24 Ohm
70 °C: 127.07 Ohm

If the temperature fluctuates strongly, you should connect a Pt100 sensor to terminals 5 + 6.

Can pH sensors also be used in drinking water?

Ja
However, only up to a temperature of <60°C.

Are the pH and redox sensors also suitable for Ex zone 1?

Ja.

How often should a pH sensor be calibrated?

Ideally, the pH measuring loop, consisting of sensor, cable and connector as well as evaluation unit, should be checked once a week with a pH buffer solution and readjusted if necessary.
This is particularly easy, because removal is greatly simplified, with the sliding retractable fitting !

Which pH sensors are suitable for deionised water?

A glass electrode (e.g. pH 110 TE) should always be used.
The pH-62 plastic electrode supplied with the PM 2000 hand-held meter is not suitable.
Conductivity and redox measurements can also be carried out in deionised water without any problems.

In which liquid must pH sensors be stored?

To ensure the longest possible storage life, the cap placed on the tip of the electrode should be filled with a KCl solution; under no circumstances should water be used.

How and for how long can pH and redox sensors be stored?

pH and redox sensors must only be stored moist. For this purpose, a protective cap with a 3-molar KCl solution is fitted to sensors in their original packaging. You can obtain KCl solution from us.

pH and redox sensors have a limited shelf life, which is why stocking for longer than 3 months(!) is not recommended. Please always store the sensors upright, not lying down; corresponding arrows are printed on the outer packaging.

What output signal is present at the cable ends at pH 7?

0 mV at pH 7, values below / above with a slope of 57 mV/pH. At pH 9, a value of -116 mV should then be measured (or reading at pH 7 - 116 mV).

Calibrating the sensor has recently been taking a very long time. What is the cause?

2 possible causes:

  • The electrode is worn out and must be replaced. Possible from 6 months of use, but certainly after an operating period of 12 months or more.
  • The electrode is simply more heavily contaminated. Clean the electrode tip with clear water. Dilute hydrochloric acid (10%) can be used to clean off lime deposits.
The sensor can no longer be calibrated. The messages displayed after calibration indicate an error.

Possible causes:

  • The sensor is defective, e.g. glass breakage.
  • The cable is defective, no longer high-impedance.
  • The instrument is defective.

If you are using a SMART pH, switch the input to mV measurement in configuration level 1. This allows you to see what voltage the pH electrode delivers in different buffer solutions via the cable to the measuring input. With buffer pH 7, 0 mV is ideal, <50 mV still OK. With buffer pH 9, ([reading of buffer pH 7] - 116 mV) is ideal, ([reading of buffer pH 7] - 100 mV) still OK.
Reset the redox offset to zero. To do this, create a short circuit (wire jumper) directly at the SMART electrode connection, removing the electrode cable from the SMART. While the input is short-circuited, start the redox calibration and enter 0000 mV there. Then remove the wire jumper, connect the cable and electrode and carry out the following tests:

  • Immerse the sensor in buffer solution pH 7 and read the mV value. Rinse the sensor and immerse it in pH 9. 0 mV and -116 mV respectively should be displayed. If so: everything is OK. If not, continue with test 2.
  • Disconnect the sensor from the cable and create a short circuit between inner conductor and shield on the cable with a small wire jumper. The SMART must now display 0 mV, (+/-5 mV) still OK. If so: the electrode is defective. If not: continue with test 3.
  • Disconnect the cable from the instrument. Short-circuit the measuring input with a wire jumper. The instrument must display 0 mV, (+/-5 mV) still OK. If so: the cable is defective. If not: the instrument is defective.
How can I measure the pH value of soil?

For measuring the pH of soil, mix about 100 cm3 of soil with 100 cm3 of distilled water thoroughly. Immerse the pH combination electrode and move it slightly. You can read the pH value when the display no longer changes. After measuring, clean the pH combination electrode only by rinsing with tap water and shake off any remaining water. After use, please put the wetting cap filled with KCl solution back over the electrode tip. Renew the KCl solution frequently.

What do I need to consider regarding cables and connectors?

Make sure that the plug connections are kept absolutely clean and dry, otherwise leakage currents may occur that reduce the measuring signal or cause it to collapse completely.

When stripping the coaxial cable, the black, semi-conductive plastic layer must be completely removed. The insulation between inner conductor and shield must not be touched with the fingers. Further information and an example here...

A standard 50 ohm coax cable is built for high-frequency applications. pH measurement involves a very high-impedance DC voltage signal. The stated characteristic impedance of 50 ohms is therefore irrelevant for pH measurement. The cable structure and material of a 50 ohm cable are not suitable for pH measurement. The risk of difficulties increases with cable length. A special pH cable should be used, which has an additional insulating layer between inner conductor and shield over the sheath of the inner conductor.

What cable lengths are possible?

When using a suitable coax cable, cable lengths of 50 m and more are possible.

When laying, make sure that no power cables run parallel or cross.

Can pH and redox sensors also be installed horizontally?

No. The sensors must be installed vertically up to max. +25°.

However, we can offer special probes for horizontal installation. Please ask us about them.

How do I calibrate a pH sensor operated on a pH transmitter?

You need 2 pH buffer solutions, pH 7 and pH 4.

In the following example we use a transmitter with 0-10 V output; the results for transmitters with 0-5 V, 0-20 mA or 4-20 mA output are given in brackets.

  1. Rinse the pH electrode thoroughly.
  2. Immerse the electrode in the pH 7 buffer solution.
  3. Use the "zero point shift" adjusting screw to set the zero point so that 5 V (2.5 V / 10 mA / 12 mA) is present at the output.
  4. Rinse the pH electrode thoroughly again.
  5. Immerse the electrode in the second buffer solution, pH 4.
  6. Use the "slope" adjusting screw to set the desired slope. If 2 V (1 V / 4 mA / 7.2 mA) is then present at the output, the slope is 1 V/pH (0.5 V/pH / 2 mA/pH / 1.6 mA/pH).
  7. The output voltage / output current at pH 4 / pH 9 can be found in this Excel file. You can also change the measuring range.
Directly after calibration, the reading deviates from the correct value by 0.3 pH. Is the electrode defective?

A pH sensor is operated on the SMART 96-2 pH and is calibrated with 2 buffer solutions, pH 4 and pH 7. After calibration, as verification, the pH sensor is immersed in both buffer solutions and the readings are taken. pH 7 is displayed correctly, but 3.7 is displayed in the pH 4 solution.

This is probably due to the configuration of the SMART, not the sensor. The SMART has the two settings "Temperature pH calibration" and "Temperature pH measurement". Different temperatures are probably set here. When the instrument then switches from calibration mode to measuring mode, temperature compensation is carried out. This explains why the zero point (i.e. pH 7) is measured correctly but the pH 4 value is corrected.

How can I check the signal output of the transmitter?

For troubleshooting on the output side, it is advisable to first disconnect the transmitter from the PLC.

Supply the transmitter with voltage via PINs 1 and 3. Then hold the ammeter to PINs 2 and 3 and read the current.

To the products: pH and redox sensors