Maintenance and calibration of pH sensors

Shipping and storage

pH and redox sensors must be stored moist. To do this, pour a little 3-molar KCl solution into the protective cap and slide it onto the sensor.

Caution!
Do not store the sensor in distilled water, as this can lead to premature ageing and defects in the reference system.

All electrodes are individually tested and supplied individually packed. They should be stored in dry rooms at 10-30°C.

At temperatures below -5°C, the electrodes may burst due to freezing of the inner buffer and inner electrolyte. When forwarding, make sure the packaging is frost-proof!

Electrodes have a limited shelf life. It is therefore advisable not to exceed a storage period of approx. 3 months.

Visually inspect the probes for trapped air bubbles. If air bubbles are present, they can be removed by shaking downwards (as with a clinical thermometer).

Calibration and checking

The cap must be removed for calibration and measurement. pH sensors stored dry should be soaked for 24 hours before use, otherwise the zero point will drift.

Every newly commissioned pH sensor must be calibrated; a 2-point calibration is recommended and standard buffer solutions pH 7 and pH 4 or pH 9 should be used:

  1. Immerse the pH sensor in buffer solution pH 7. (With automatic temperature compensation, the temperature sensor must also be immersed in the buffer solution.)
  2. Use the "zero point" adjusting knob to set the display on the instrument to the value of the buffer solution.
  3. Rinse the sensor with distilled water
  4. Immerse the sensor in the second buffer solution, e.g. pH 4.
  5. Use the "slope" (mV / pH) adjusting knob to set the display on the instrument to the value of the second buffer solution.

The pH measuring device is now calibrated. The frequency of calibration or cleaning of the sensor depends on the operating conditions (contamination, chemical load).

Redox sensors are not calibrated. A check should merely be carried out at regular intervals using a redox reference solution.

Cleaning

The sensor must be cleaned before each calibration. Experience shows that the following agents can be used for cleaning:

  • For general deposits: non-abrasive household cleaners
  • For greasy, oily contamination: surfactant-containing cleaners (e.g. washing-up liquid or acetone).
  • For lime deposits and metal hydroxide coatings: dilute hydrochloric acid (10%)
  • For sulphide-containing deposits (e.g. in sewage treatment plants): cleaning mixture of dilute hydrochloric acid (10%) and thiourea (saturated)
  • For protein-containing media (measurements in the food sector): cleaning mixture of dilute hydrochloric acid (10%) and pepsin (saturated)

In principle, sufficient rinsing is required after each cleaning.

Before reuse, the pH sensor must be soaked in saturated KCl solution for approx. 2 hours.

Caution!
Observe safety regulations when handling acidic solutions. After cleaning, always rinse all parts with deionised water.

The metal surface of redox sensors can also be cleaned by grinding and polishing. If the ceramic diaphragm of the reference system mounted on the side is blocked, it can be cleaned like the glass membrane and additionally by carefully scraping with a fingernail, a razor blade or a fine file. Take great care not to scratch the glass membrane.

Installation

Redox and pH sensors must be installed vertically from above – with a max. angle of 80° to the vertical. Sensors installed horizontally or upside down cannot measure.

Service life

Redox and pH sensors are subject to natural ageing even when handled properly. Depending on the application, a service life between half a year and max. three years can be stated. In individual cases, especially under extreme operating conditions, the service life may be reduced to days.