Is measurement without a Pt100 possible with the SMART?

Ja.
As factory setting, configuration level 1 has the setting temperature=AUS selected, i.e. operation without a temperature sensor.
However, automatic temperature compensation is then not possible.
On the SMART LF, set temperature compensation in configuration level 5 toAUS ormanual. If you select manual, also set the reference temperature. We recommend the settings temperature compensation=manual and reference temperature=25°C.
Temperature has less effect on pH measurement. In configuration level 1, set temperature=AUS and in configuration level 5, under temperature compensation pH measurement, select the temperature of the medium
If you have connected a Pt100 to the SMART, set temperature in configuration level 1 toEIN and temperature compensation in configuration level 5 toAUTO.

Which conductivity sensors can be operated on the SMART LF?

You can connect the 2-electrode sensors of types LMP (= with integrated Pt100) and LM (without temperature sensor) or the 4-electrode sensors LMP1-4E-PP or LMP5-4E-PP to the SMART LF instruments.
If an LM-type sensor is used, the type of temperature compensation must be set in configuration level 5.

Can a dosing pump, for example, be controlled directly with the SMART?

Ja.
Common pumps have a signal input to which the relay output of the SMART can be connected. In this case, depending on the pump used, the SMART output should be operated as a proportional pulse length controller or proportional pulse frequency controller.
If the pump does not have a signal input, it can be controlled by switching the relay output of the SMART directly into the power supply. However, to avoid voltage peaks that could damage the relay contacts in the long term, an RC element for smoothing, for example, should then be provided.

Can the SMART WG instruments be used in Ex zone 1?

Unfortunately not
The SMART WG instruments have not been tested for operation in this area, as the number of enquiries would not justify the cost. The opening in the housing for the display is sealed by the membrane keypad above it, as this has been bonded all round. The cable glands also fit tightly; however, this depends on the type of installation (e.g. sealing cannot be guaranteed if the cables used are too thin). The decision as to whether the instrument can be used in a protected area must therefore be made by the user!

How do I set the analogue output of the SMART instruments?
  • In configuration level 1, under the menu item Input, select the desired measured variable (pH for pH measurement, mV for redox measurement).
  • In configuration level 1, under the menu item Analogue 1 (or 2), set the current output: (0…20 mA or 4…20 mA) and set whether 20 mA is the maximum or the minimum.
  • Then be sure to set the measuring range in configuration level 2! Under the menu item Analogue 1 (or 2), first set the upper limit of the measuring range (arrow up on the display) and then the lower limit.