FAQ: Redox measurement

What is redox?

The term redox is an abbreviation for the simultaneous reaction of reduction and oxidation. Oxidation means not only the uptake of oxygen but, more generally, the uptake of positive charges or release of negative charges. Reduction means the reversal of such processes.

How is the redox potential measured?

For this you need a platinum electrode and a pH meter with mV measuring range. The platinum electrode is immersed in the water like the pH electrode and the measurement result is read after some time.

Electrode, measuring chain or sensor. Which is correct?

The terms electrode, measuring chain or combination electrode are often used synonymously.
Usually, combination electrodes are meant, i.e. an electrode combined with a reference electrode, whereby the reference electrode also refers to the reference system.
We consistently use the term Sensor.

What exactly is a combination electrode?

Measuring chains for measuring electrochemical potentials always consist of a measuring electrode and a reference electrode. The measuring electrode outputs a voltage in relation to the concentration. The voltage of the reference electrode, on the other hand, is constant and is not influenced by the medium.
Combination electrodes form a single unit of glass and reference electrode. The reference electrode surrounds the glass electrode concentrically.

Reference electrode system:
Reference electrode systems are mainly equipped with a silver/silver chloride reference system. Reference systems with liquid electrolyte should always contain sufficient liquid; if necessary, they are topped up with 3 mol potassium chloride solution. During measurement, the filling opening should be exposed for pressure equalisation. The liquid column inside the electrode should rise a few centimetres above the surface of the measuring solution to create a hydrostatic overpressure, which ensures a constant flow of KCl solution through the diaphragm out into the measuring solution. This ensures constant self-cleaning of the diaphragm and also prevents possible electrode poisons from diffusing into the electrode and rendering the lead-off system unusable. Reference electrodes with gelled reference electrolyte cannot be refilled.
Diaphragms:
Reference electrodes are connected to the measuring solution via a diaphragm. This diaphragm is porous and permeable to liquid and usually consists of sintered ceramic.
Can redox sensors also be used in drinking water?

Ja
However, only up to a temperature of <60°C.

Are the redox sensors also suitable for Ex zone 1?

Ja.

How often should a redox sensor be calibrated?

A redox sensor cannot be calibrated. However, check regularly with a reference solution whether the sensor is still working properly:
Connect the sensor to the controller. Immerse the sensor in a redox buffer solution. The value of the buffer solution should be reached or exceeded within max. 30 sec. If the sensor reacts very slowly or the reading falls more than 20 mV below the value of the buffer solution, the sensor must be cleaned. If this does not help either, the sensor must be replaced.

What do I need to consider regarding cables and connectors?

Make sure that the plug connections are kept absolutely clean and dry, otherwise leakage currents may occur that reduce the measuring signal or cause it to collapse completely.

When stripping the coaxial cable, the black, semi-conductive plastic layer must be completely removed. The insulation between inner conductor and shield must not be touched with the fingers. Further information and an example here...

How do I use the zero point shift of the redox transmitter?

You need the redox solution UH=468mV.

In the following example we use a transmitter with 0-10 V output; the results for transmitters with 0-5 V, 0-20 mA or 4-20 mA output are given in brackets.

A redox electrode cannot be calibrated, so the slope cannot be changed. However, you can shift the zero point on the transmitter.

On delivery, the zero point is set so that 5 V (2.5 V / 10 mA / 12 mA) is present at the output with the input bridged.

  1. Rinse the redox electrode thoroughly.
  2. Immerse the electrode in the redox solution.
  3. Use the "zero point shift" adjusting screw to set the zero point. 6.95 V (3.475 V / 13.9 mA / 15.12 mA) is then present at the output.

To the products: pH and redox sensors