FAQ: Conductivity measurement

What does conductance mean?

Conductance, or conductivity, refers to specific conductivity. Liquids conduct electric current by means of ions. The more ions, the higher the conductivity. Units: µS/cm (microsiemens per centimetre) or mS/cm (millisiemens per centimetre).

What is conductivity?

Most of the ions in water are parts of dissolved salts. For example, common salt (chem. NaCl) dissociates into a positive sodium ion (Na+) and a negative chloride ion (Cl-). Both types of ion conduct electric current. Put simply, conductivity is a measure of the amount of salts dissolved in the water.

What does the cell constant mean, e.g. CC0.5, CC1, ...?

Conductivity sensors consist of a pair of electrodes. The cell constant K is an essential property of the sensor and reflects the ratio of electrode spacing to electrode area.
Because the conductivity between the electrodes is influenced by the cell geometry, area and spacing must be defined as the cell constant.
For a traditional cell with two platinum electrodes of 1 cm² area at a distance of 1 cm, the cell constant is K = 1.0 cmˉ¹.

  • C = conductivity in siemens per cm (S/cm)
  • R = electrical resistance in ohms
  • G = conductance in siemens (S; 1 S = 1 ohmˉ¹)
  • L = electrode spacing (cm)
  • A = electrode area (cm²)
  • K = cell constant (cmˉ¹); K = L/A

For weakly conductive samples, the electrodes are moved closer together by reducing L in order to obtain smaller cell constants of 0.1 and 0.01 cm-1. This increases the conductance between the electrodes and thus simplifies electronic evaluation by the instrument. Conversely, L is increased for highly conductive solutions (cell constant K = 3 cmˉ¹)

What must be considered regarding temperature compensation with conductivity sensors?

If LMP sensors are used, the temperature compensation is calculated in the evaluation unit (e.g. SMART, but not ISO transmitters). This has an extra input for the Pt100 signal and compensates (a) linearly or (b) naturally, in the range 0…2.2% per kelvin according to a fixed mathematical formula relating heat and conductance. Most users use natural compensation. The curve for b) is only approximately linear, rather banana-shaped.

If LMN sensors are used, the temperature compensation is linear in the ISO transmitter. Unlike the SMART, this cannot perform any calculations as it has no CPU, i.e. a Pt100 cannot be evaluated.

What signal output do the conductivity sensors have?

All conductivity sensors are passive / conductive, i.e. they must be connected to an active device (e.g. transmitter or SMART). This then provides usable current or voltage signals.

I would like to operate a conductivity sensor with cell constant 1 up to max. 200 mS. Is that possible?

High mS values often go hand in hand with sludge-laden media; therefore, where possible, a conductivity cell less susceptible to contamination should be preferred here, i.e. type 9 (CC-1) rather than type 6 (CC-5).
Although optimally operated only up to 40 mS, CC-1 cells can be used e.g. on the SMART 96-2 LF up to max. 200 mS!

What different designs do your conductivity sensors have?

The design of the conductivity sensors depends exclusively on the cell constant, i.e. it has nothing to do with the application.
A typical question is, for example, whether the cone of the CC-5 cells would also be suitable for flow measurement in a pipe. It is, since the design has nothing to do with the field of application.

ZK 0.01 + ZK 0.1: small circular opening
ZK 1: large circular opening
ZK 5: Cone
How can the conductivity of heavy fuel oil be measured?

Under no circumstances with the conductive sensors offered by IMP GmbH.
With these, whether fitted with graphite or stainless steel plates, the surface of the plates must remain clean. The heavy fuel oil would coat the stainless steel surface or penetrate the graphite and make measurement impossible.
Sensors based on the inductive measuring method may be suitable.

What advantages do I have when using a 4-electrode sensor?

Contamination of the electrode with insulating substances (oils etc.) changes the cell constant and falsifies the measurement results.
With the 4E measuring cell, a currentless electric field measurement is therefore additionally performed via sense measurement. The 4E electrode is therefore less sensitive to contamination.
The evaluation unit must support 4-electrode measuring cells (e.g. SMART 96-2 LF). The 4E measuring cells are available in a CC1 and CC5 variant, whereby the cell constant can only be stated approximately in advance and, for example, may be between 0.85 and 1.15 for the CC1 variant. The label on the measuring cell then states the exact cell constant value; this must then be entered on the display unit.

What output current can the optocoupler of the LMLIU be loaded with?

With max. 5 mA

Do conductivity sensors have to be calibrated?

No, calibration is neither possible nor necessary. However, the electrodes should be cleaned regularly.

How are conductivity sensors cleaned?

The electrodes should be cleaned regularly.

Stainless steel electrodes (e.g. LM-LIU) can be cleaned in 3 ways:

  • For whitish or brown deposits, place the electrodes in 3% hydrochloric acid for a few minutes until the deposits dissolve. Then rinse (absolutely and very thoroughly!) with clear water (DI water if necessary)
  • For organic contamination, i.e. biofilm or oil contamination, immerse the electrodes in an ultrasonic bath with a degreasing additive (e.g. washing-up liquid) or
  • clean the electrodes with alcohol, solvents or soapy water. In addition, you can remove the contamination by mechanically cleaning the inner pin and outer electrode with a soft brush. However, these are not ideally accessible. Make sure that the surface of the electrodes is not damaged during mechanical cleaning, otherwise the cell constant will change.

Carefully clean graphite electrodes with a soft brush and a little scouring cream. Then rinse with clear water. After cleaning, the electrodes should be well wetted by the water, i.e. the water should not bead off. Alternatively, graphite electrodes can also be cleaned in 3% hydrochloric acid. Here too: then rinse extremely thoroughly with clear water. Make sure that the surface of the electrodes is not damaged during mechanical cleaning, otherwise the cell constant will change.

Up to what medium temperature can the LMLIU be operated?

Up to approx. 50°C, regardless of the housing material (PP or PVDF).
The electronics are located so close to the electrode and housing that excessive heating of the analogue and digital components is to be expected at temperatures above 50°C.

Can the Pt100 sensor of an LMP measuring cell be used for temperature measurement?

If temperature compensation is not required or desired, the Pt100 sensor can be used for temperature measurement. The sensor is somewhat slower than a dedicated Pt100 sensor. It is connected to a Pt100 input of the display unit using the two connecting leads assigned to the sensor.

Is it advisable to operate conductivity sensors beyond the recommended measuring range?

Example: a sensor with cell constant CC=0.1 and a recommended measuring range up to 500 µS is to be used on the transmitter up to 2000 µS. This is possible with the corresponding setting on the transmitter. However, the cell should only be used for measurements up to 2000 µS if measurement in the high range beyond the recommended range is only short-term and does not need to be very exact – i.e. if measurements are mainly below 500 µS but the value may rise briefly. For exact, continuous measurements beyond the recommended measuring range, a cell with the next higher cell constant should be used.

What is the maximum cable length that can be used with conductivity sensors?

The conductivity cable can also be 50 - 100 metres long. Conductivity measurement does not become slower if the cable is longer; however, if you are in a low measuring range (e.g. 50 µS), the measurement is more sensitive than if you are at e.g. 20 mS.

How is the total measurement uncertainty calculated for a measuring chain consisting of transmitter and measuring cell?

The tolerance of the cell constant of the measuring cell is ±3% for CC=1. While this tolerance can be compensated on the SMART display and control units by setting the cell constant accurate to the second decimal place, it must be included in the consideration of measurement inaccuracy for the measuring chain of measuring cell and transmitter.

The transmitter has a measurement uncertainty of ±1% (for measurements up to 2 mS) of the currently measured value (i.e. not of the full-scale value!). Only with very small readings is the measurement uncertainty somewhat more noticeable, since the 1% uncertainty then lies very much in the decimal range.

The measurement uncertainty is therefore calculated from the tolerance of the cell constant and the measurement uncertainty of the transmitter.

Which thread types are fitted to the LM/LMN/LMP/LMLIU/LMA05 conductivity sensors?

The 3/4" threads are type G (cylindrical), the 1/2" threads type NPT (tapered).

The protective cap of the LMA05 with CC=4 is very difficult to pull off for cleaning the electrodes. Can't I simply leave it off?

The protective cap does indeed sit very tightly and may have to be clamped to pull it off, e.g. in a vice. Then simply pull out the probe body.
However, the protective cap must always be refitted after cleaning the electrodes, as it significantly influences the cell constant of the sensor.
So the answer to your question is no, the protective cap must never be left off.

Can LM/LMN/LMP sensors be operated in alkaline liquids (pH 14) and at temperatures of 70°C?

We are not aware of any problems so far when operating the sensor in alkaline media. The temperature range stated in the data sheet must be observed. This is 70°C for PP.

Despite changing media, the reading of my LM-type conductivity sensor remains constant at 5 (or 50 or 500, depending on the set cell constant).

The measuring input of the evaluation unit (SMART or transmitter) is probably connected to terminals 1 and 3 of the conductivity sensor, but must be connected to terminals 2 and 4 (ground).

Background: LM-type sensors have a fixed 2 kOhm resistor installed instead of a temperature sensor. This resistor is connected to terminals 1 and 3, and its reciprocal is displayed as a constant conductance if the measuring input of the SMART or transmitter is connected to these two terminals.

With an LMN-type sensor the error also occurs with the constant value 5, 50 or 500; with an LMP-type sensor, a constant value of 10, 100 or 1000 is displayed if connected incorrectly.

How can I check the signal output of the transmitter?

For troubleshooting on the output side, it is advisable to first disconnect the transmitter from the PLC.

Supply the transmitter with voltage via PINs 1 and 3. Then hold the ammeter to PINs 2 and 3 and read the current.

To generate the minimum value, simply take the conductivity sensor out of the medium.

To generate the maximum value, you can bridge the electrodes of the conductivity sensor with a wire.

Can LM/LMN/LMP sensors be used in hydrochloric acid with a concentration of 15%-20%?

Unfortunately not. Inductive sensors are preferable here.